redtel-blutdruck.de/en/knowledge/validation

How is a measuring method tested scientifically?

Answer in 20 seconds

By holding it against something already known — the reference. For blood pressure that is the invasively measured pressure in the middle of the aorta. Only when the new method reproduces the same course as the reference, and not only at rest, is the foundation laid for calibration and approval.

5 min read Updated 4 September 2026 8 articles · 0 videos · 1 publication

The non-invasively recorded pressure course compared with the invasive reference measurement in the aorta.
From the article “Why must the BPER match the invasive reference?”
Explanation

The test begins at the reference, not at the sensor

Continuous blood pressure measurement for wearables is usually presented as a sensor question: a better component, better optics, a better result. But the difficulty starts earlier.

Before anything can be calibrated, it has to be shown that a non-invasive method captures the same pressure course as the invasive measurement. Only when both see the same changes is the data usable for a calibration.

The central question is therefore not which sensor is used. It is whether the non-invasive measurement reproduces the same physiological reality as the reference.

Explanation

Why the site of the reference decides

Even an invasive measurement is not equally informative everywhere. Close to the vessel wall, local influences take effect: wall tension, reflections of the pressure wave, haemodynamic effects at the measuring point.

Measuring in the middle of the aorta, by contrast, gives the central reference pressure and removes a large part of these error sources. Testing against a badly placed reference produces deviations that say nothing about the method being tested.

For professionals

The steps of a validation

1 — Define the reference
What counts as true? Site, method and conditions have to be defined beforehand.
2 — Measure simultaneously
Both methods on the same person at the same moment. One after the other is not a comparison.
3 — Compare the course
Not only single values, but whether the same changes become visible at the same time.
4 — Test under change
Load, medication, change of posture, illness. Limits are defined by change — so that is where testing has to happen.
5 — Disclose
Describe method, data and limits so that others can repeat the test.

At rest almost every method agrees. Evidence that covers only the resting state therefore says little about everyday life.

REDTEL position

What was tested in Paper 1

The BPER rests on the second-order blood pressure fluctuation — not on a static single value, but on the breath-driven dynamics of the circulation, which look different in every person.

It is exactly this foundation that was tested, for the first time, directly against invasive aortic pressure. That is the content of Paper 1: not another calibration method, but the precondition for stress-free blood pressure measurement to work correctly at all.

Calibration here is not the blood pressure value itself. It is the demonstrated relationship between a measured signal and the actual pressure. The data from wearables stays the same — the open question is how the calibration is done.